In use test (phase 3 test) of Channel Cleaner Ball Brushes applied for manual cleaning of flexible endoscopes
|
|
- Steffen Bjerre
- 8 år siden
- Visninger:
Transkript
1 In use test (phase 3 test) of Channel Cleaner Ball Brushes applied for manual cleaning of flexible endoscopes Endoscopy Unit Surgical Gastrointestinal Department K Bispebjerg Hospital Copenhagen University Hospital Clean-endoscope Hvidovre Hospital Copenhagen University Hospital Denmark
2 In use test (phase 3 test) of Channel Cleaner Ball Brushes applied for manual cleaning of flexible endoscopes Introduction Manual cleaning and brushing of the channels in a flexible endoscope (FE) are very important before disinfection. If debris, blood or mucus membranes are remained in the channels, the disinfection efficacy may get negatively affected 1. The quality of hand manual cleaning and brushing can vary between operators. The Channel Cleaner ball brush was designed and developed to supplement manual cleaning and brushing. Experimental studies showed that the product is effective in lowering the protein content in Teflon tubes artificially contaminated in comparison with standard brushes 2. ATP bioluminescence is used for measuring levels of organic soil on surfaces. Opposed to a Microbiologic control samples an ATP cleaning test can be performed in less than 2 minutes. Residual soil is an indicator that the surfaces may be unclean and could provide a potential reservoir and harbour for bacteria, fungi and virus increasing the risk of cross infection between patients 3. Aim of the study Aim of the study is to compare the manual cleaning of patient-used gastroscopes and colonoscopes performed either with Channel Cleaner Ball Brushes or a standard brush. Product: Channel Cleaner Ball Brushes (together with Cleantop-21 detergent and UNO CLEAN Valvebrush type AN ) Order from Braun Scandinavia A/S Lervej DK-3500 Værløse Washer disinfectors ETD2+ and ETD 3 with per acetic acid (PAA) Chemical products For manual cleaning: CLEANTOP-21 2
3 Cleaner in WD: Disinfectant in WD: EndoDet 0.6%, Ecolab EndoDis 1.2% + EndoAct 1.2%, Ecolab Endoscopy department The Endoscopy Unit Surgical Gastrointestinal Department K Bispebjerg Hospital Copenhagen University Hospital Flexible endoscopes included Routine gastroscopes and colonoscopes Test laboratory Clinical Microbiology Department, Hvidovre Hospital, Copenhagen University Hospital, has trained the endoscopy staff at. Condition for the investigation The endoscopy unit follows the Danish recommendations for cleaning and disinfection of FEs 4. In the reprocessing area manual cleaning is performed of all channels and the outer of the FE in a detergent solution. Suction- and biopsy channels are brushed. The detergent solution is discarded after each use. Before storage we flush the channels with 70% alcohol, and the FEs can be used within 3 days. ATP-test Residual Adenosine Tri-Phosphate (ATP), which is found in large quantities in human blood and other tissue fluids, are used as cleanness indicator for FE channels. Low-level ATP residuals indicate that patient related secretes are removed (no hepatitis B + C risk), and that the subsequent disinfection will be successfully. Log CFU (CFU = colony forming units) and Log ATP are not correlated on surfaces contaminated with human secretes. 3
4 An UNI-LITE NGl portable luminometer and total ATP Aqua-Trace were used (3M). The relation between Relative Light Units (RLU) values from UNI-LITE NGI ATP are earlier computed as log RLU = Log ATP + 1 (femtomol), R 2 = Log RLU = Log ATP-units (10-14 mol). 3 log ATP-units 10-3 μl blood. Pre-cleaning The suction and biopsy channel of the FE were flushed with a 1% cleantop-21. The water/air channels were flushed with water with the help of a dedicated valve. The insertion section was wiped with the detergent solution. Pre cleaning sampling The biopsy valve was replaced with a stop. Water from a 20 ml syringe was injected through the suction port on the light-guide connector and collected in a 30 ml centrifuge container at the tip of the insertion tube. The was processed according to manufacturer s instruction. Manual cleaning Brushing Period 1 and 3: Manual cleaning with a standard brush of adequate size. Period 2 and 4: 3 Ball Brushes are sucking through the biopsy/suction channel. Biopsy- air/water- and suction valves were removed. All inlets were cleaned with UNO CLEAN Valvebrush type The full FE was immersed and wiped in the detergent solution (Cleantop-21). The endoscope was removed from the detergent solution and stops were inserted in suction valve and biopsy inlet. Post cleaning sampling Water from a 20 ml syringe was injected through the suction port on the light-guide connector and collected in a 30 ml centrifuge container at the tip of the insertion tube. The was processed according to manufacturer s instruction. 4
5 Study design ATP sampling from biopsy/suction channel Brush/balls Sampling Sample size Brush Before manual cleaning Balls Brush Balls After manual cleaning One week without sampling Before manual cleaning After manual cleaning One week without sampling Before manual cleaning After manual cleaning One week without sampling Before manual cleaning After manual cleaning Data presentations 1. Equivalence chart: x = log ATP-units pre cleaning; y = log ATP-units post cleaning. On the equivalence line ATP residuals are the same before and after the manual cleaning. 2. Results were entered in Six Sigma Control charts with determination of centreline (CL), upper control limit (UCL) = CL + 3Σ and lower control limit (LCL) = CL - 3Σ. CL (mean) represents the central tendency of the cleaning effect 5. a. Limits: All results should be inside the control limits. i. The upper limit suggested of Obee et al is 2.7 log ATP-units 3 5
6 Log ATP-units Results Before manual cleaning: The mean of log ATP-units was 4.2 and 3.8 after the pre-cleaning of the biopsy/suction channels of the FEs in the brushes and the Ball Brushes periods respectively. Figure 1 shows a ATP level are within the control limits. Therefore the pre-cleaning process in the endoscopy rooms are a statistical stable processes, but with different means and control limits in the brush and Ball Brush periods. ATP level before manual cleaning of the biopsy/suchtion channel with a cleaning brush or Channel Cleaner Ball Brushes + UCL 7 6 UCL 5 Mean 4 Mean 3 2 LCL 1 LCL 1 6 Sample Number (1-50 brushes; Ball Brushes) Figure 1: ATP level after pre-cleaning before manual cleaning of the biopsy/suction channel of flexible endoscopes with either a brush or Ball Brushes. UCL: Upper control limit. LCL: Lower control limit. After manual cleaning: The means of log ATP-units were 1.9 in both the brushes (Range: ) and Ball Brushes (range: ) periods. Figure 2 shows that all results are inside the control limits and the processes are statistical stable. 96 % and 100 % of the FEs passed the upper limit suggested of Obee et al (2.7 log ATP-units) in the brush and Ball Brushes periods respectively. The upper limit of Obee at al. seems to be a mean + 2 Sigma limit. Figure 3 shows an effective cleaning process with only one control points above the equivalence line. All other control points are below the equivalence line. There is no significant correlation between log ATP-units before and after the manual cleaning. 6
7 Log ATP-units ATP residuals after manual cleaning of the biopsy/suchtion channel with a cleaning brush or Channel Cleaner Ball Brushes * UCL 3,0 Obee et al: Upper limit 2,5 2,0 Mean 1,5 1,0 LCL 1 6 Sample number (1-50 brushes, Ball Brushes) Figure 2: ATP residuals after manual cleaning of the biopsy/suction channel of flexible endoscopes with either a brush or Ball Brushes. UCL: Upper control limit. LCL: Lower control limit. Figure 3: Relation between ATP residuals before and after manual cleaning of the Biopsy/suction channel of flexible endoscopes with a brush or Ball Brushes. 7
8 Conclusion Manual cleaning with Channel Cleaner Ball Brush 100% of the biopsy/suction channel of flexible endoscopes passed the upper limit suggested of Obee et al (2.7 log ATP-units) after 50 manual cleaning procedures performed with Channel Cleaner Ball Brush and a Cleantop-21 detergent. The cleaning process was in statistical control (all control points inside the control limits in a Six Sigma control chart) The range of log ATPunits was Manual cleaning with a normal cleaning brush 96% of the biopsy/suction channel passed the test after 50 manual cleaning procedures with single use cleaning brushes. The cleaning process was in statistical control. The log ATP-unit difference between brush and Ball Brushes was not significantly. The log ATP-unit range was The range in the brush and the Ball brushes periods were lower than the range at log ATPunits accounted by Obee et al 3 Channel Cleaner Ball Brush can on the above be recommended for manual cleaning of flexible endoscopes in combination with Cleantop-21 and careful cleaning of all inlets with the UNO CLEAN Valvebrush type AN Torsten Slotsbjerg (Torsten Slotsbjerg) Consultant Clean-endoscope Clinical microbiologic department Hvidovre Hospital Denmark 8
9 Klinisk brugstest af Channel Cleaner Ball Brushes Dansk resume Channel Cleaner Ball Brushes (fiberkugle børster) er udviklet til anvendelse ved manuel rengøring og børstning af fleksible endoskoper. Laboratorieundersøgelser har vist, at produktet er mere effektivt til at fjerne proteinrester fra teflonrør end de standard børster, som normalt anvendendes ved manuel rengøring af endoskoper. 2. ATP bioluminiscens anvendes til at påvise mængden af organisk materiale på overflader. Adenosin Trifosphat (ATP) findes især i store mængder i blod, spyt, urin og andre friske udskillelser fra mennesker (og dyr). I modsætning til en proteintest kan en ATP rengøringstest udføres på mindre end 2 min. Et UNI-LITE NGl transportabelt luminometer og et total ATP Aqua-Trace testkit er anvendt ved ATP-bestemmelsen (3M). Design af studiet ATP indholdet i biopsi- og sugekanalen på gastro- og koloskoper er bestemt før og efter manuel rengøring i perioder enten med traditionelle engangs rengøringsbørster eller med Channel Cleaner ball brushes. Der er foretaget overkrydsning af perioderne efter nedenstående skema. Som detergent er anvendt Cleantop-21, et enzymfrit detergent. Børster/kugler Prøvetagning Antal prøver Børster Før manual rengøring Kugler Efter manual rengøring En uge uden prøvetagning Før manual rengøring Børster Efter manual rengøring En uge uden prøvetagning Før manual rengøring Kugler Efter manual rengøring En uge uden prøvetagning Før manual rengøring Efter manual rengøring 9
10 Kritiske grænser Obee og medarbejdere har forslået en øvre grænse for ATP svarende til 2.7 log ATP-enheder. Resultaterne af ATP målingerne indføres i 6-Sigma kontrolkort og alle resultater skal ligge indenfor kortets kontrolgrænser. Resultater. Channel Cleaner Ball Brushes perioderne 100 % af biopsi/sugekanalerner opfyldte de opstillede krav, efter 50 manuelle rengøringsprocedurer af endoskoperne. Rengøringsproceduren var statistisk stabil (alle resultater inden for kontrolgrænserne i kontrolkortet). Traditionel børste perioderne 96 % af biopsi/sugekanalerner opfyldte de opstillede krav efter 50 manuelle rengøringsprocedurer af endoskoperne. Rengøringsproceduren var statistisk stabil. Forskellen i resultaterne efter den manuelle rengøring i traditionel børste og Channel Cleaner ball brush perioderne er ikke statistisk signifikant. Afsluttende resume Channel Cleaner Ball Brushes kan sammen med Cleantop-21 anvendes i stedet for en traditionel børste ved den manuelle rengøring af fleksible endoskoper. Man skal dog huske at anvende en ventilbørste til rengøring af alle indgange til endoskopets kanaler. I denne undersøgelse er der anvendt en ventilbørste af typen UNO CLEAN AN Torsten Slotsbjerg (Torsten Slotsbjerg) Consultant Clean-endoscope Clinical microbiologic department Hvidovre Hospital Denmark 10
11 Reference list 1. Chu NS, McAlister D, Antonoplos PA. Natural bioburden levels detected on flexible gastrointestinal endoscopes after clinical use and manual cleaning. Gastrointest Endosc. 1998;48: Lionel Pineau et all. An alternative to Standard Brushing Procedure. Gastrointest Endosc. 2008;67,5:AB Obee PC, Griffith CJ, Cooper RA, Cooke RP, Bennion NE, Lewis M. Real-time monitoring in managing the decontamination of flexible gastrointestinal endoscopes. Am J Infect Control. 2005;33: Råd og Anvisninger, Statens Seruminstitut. 2001;4 edition, Benneyan JC, Lloyd RC, Plsek PE. Statistical process control as a tool for research and healthcare improvement. Qual Saf Health Care. 2003;12:
Basic statistics for experimental medical researchers
Basic statistics for experimental medical researchers Sample size calculations September 15th 2016 Christian Pipper Department of public health (IFSV) Faculty of Health and Medicinal Science (SUND) E-mail:
Læs mereQuality indicators for clinical pharmacy services
Quality indicators for clinical pharmacy services Head of Quality and Improvement, Nordsjælland Hospital Dorthe Vilstrup Tomsen Assuring quality in clinical pharmacy services Following international, national
Læs mereTM4 Central Station. User Manual / brugervejledning K2070-EU. Tel Fax
TM4 Central Station User Manual / brugervejledning K2070-EU STT Condigi A/S Niels Bohrs Vej 42, Stilling 8660 Skanderborg Denmark Tel. +45 87 93 50 00 Fax. +45 87 93 50 10 info@sttcondigi.com www.sttcondigi.com
Læs mereSmall Autonomous Devices in civil Engineering. Uses and requirements. By Peter H. Møller Rambøll
Small Autonomous Devices in civil Engineering Uses and requirements By Peter H. Møller Rambøll BACKGROUND My Background 20+ years within evaluation of condition and renovation of concrete structures Last
Læs mereDIVAR VIGTIGT! / IMPORTANT! MÅL / DIMENSIONS
DIVAR VIGTIGT! / IMPORTANT! VIGTIGT læs vores anvisninger før du bruger produktet. Har du problemer med den elektriske installation, skal du kontakte en elektriker. Sørg for at altid slukke for strømmen
Læs mereFejlbeskeder i SMDB. Business Rules Fejlbesked Kommentar. Validate Business Rules. Request- ValidateRequestRegist ration (Rules :1)
Fejlbeskeder i SMDB Validate Business Rules Request- ValidateRequestRegist ration (Rules :1) Business Rules Fejlbesked Kommentar the municipality must have no more than one Kontaktforløb at a time Fejl
Læs mereMeasuring the Impact of Bicycle Marketing Messages. Thomas Krag Mobility Advice Trafikdage i Aalborg, 27.08.2013
Measuring the Impact of Bicycle Marketing Messages Thomas Krag Mobility Advice Trafikdage i Aalborg, 27.08.2013 The challenge Compare The pilot pictures The choice The survey technique Only one picture
Læs mereBusiness Rules Fejlbesked Kommentar
Fejlbeskeder i SMDB Validate Business Request- ValidateRequestRegi stration ( :1) Business Fejlbesked Kommentar the municipality must have no more than one Kontaktforløb at a time Fejl 1: Anmodning En
Læs mereFejlbeskeder i Stofmisbrugsdatabasen (SMDB)
Fejlbeskeder i Stofmisbrugsdatabasen (SMDB) Oversigt over fejlbeskeder (efter fejlnummer) ved indberetning til SMDB via webløsning og via webservices (hvor der dog kan være yderligere typer fejlbeskeder).
Læs mereStatistik for MPH: 7
Statistik for MPH: 7 3. november 2011 www.biostat.ku.dk/~pka/mph11 Attributable risk, bestemmelse af stikprøvestørrelse (Silva: 333-365, 381-383) Per Kragh Andersen 1 Fra den 6. uges statistikundervisning:
Læs mereUnitel EDI MT940 June 2010. Based on: SWIFT Standards - Category 9 MT940 Customer Statement Message (January 2004)
Unitel EDI MT940 June 2010 Based on: SWIFT Standards - Category 9 MT940 Customer Statement Message (January 2004) Contents 1. Introduction...3 2. General...3 3. Description of the MT940 message...3 3.1.
Læs mereDIVAR VIGTIGT! / IMPORTANT! MÅL / DIMENSIONS. The DIVAR wall lamp comes standard. with 2.4 m braided cord and a plug in power supply (EU or UK).
DIVAR VIGTIGT! / IMPORTANT! VIGTIGT læs vores anvisninger før du bruger produktet. Har du problemer med den elektriske installation, skal du kontakte en elektriker. Sørg for at altid slukke for strømmen
Læs mereX M Y. What is mediation? Mediation analysis an introduction. Definition
What is mediation? an introduction Ulla Hvidtfeldt Section of Social Medicine - Investigate underlying mechanisms of an association Opening the black box - Strengthen/support the main effect hypothesis
Læs mereLinear Programming ١ C H A P T E R 2
Linear Programming ١ C H A P T E R 2 Problem Formulation Problem formulation or modeling is the process of translating a verbal statement of a problem into a mathematical statement. The Guidelines of formulation
Læs mereATEX direktivet. Vedligeholdelse af ATEX certifikater mv. Steen Christensen stec@teknologisk.dk www.atexdirektivet.
ATEX direktivet Vedligeholdelse af ATEX certifikater mv. Steen Christensen stec@teknologisk.dk www.atexdirektivet.dk tlf: 7220 2693 Vedligeholdelse af Certifikater / tekniske dossier / overensstemmelseserklæringen.
Læs mereGeneralized Probit Model in Design of Dose Finding Experiments. Yuehui Wu Valerii V. Fedorov RSU, GlaxoSmithKline, US
Generalized Probit Model in Design of Dose Finding Experiments Yuehui Wu Valerii V. Fedorov RSU, GlaxoSmithKline, US Outline Motivation Generalized probit model Utility function Locally optimal designs
Læs mereUsing SL-RAT to Reduce SSOs
Using SL-RAT to Reduce SSOs Daniel R. Murphy, P.E. Lindsey L. Donbavand November 17, 2016 Presentation Outline Background Overview of Acoustic Inspection Approach Results Conclusion 2 Background Sanitary
Læs mereDumped ammunition - an environmental problem for sediment management?
5th International SedNet Conference, 27th-29th May 2008, Oslo, Norway Dumped ammunition - an environmental problem for sediment management? Jens Laugesen, Det Norske Veritas Harald Bjørnstad, Forsvarsbygg
Læs mereGusset Plate Connections in Tension
Gusset Plate Connections in Tension Jakob Schmidt Olsen BSc Thesis Department of Civil Engineering 2014 DTU Civil Engineering June 2014 i Preface This project is a BSc project credited 20 ECTS points written
Læs mereWIO200A INSTALLATIONS MANUAL Rev Dato:
WIO200A INSTALLATIONS MANUAL 111686-903 Rev. 1.01 Dato: 10.01.2013 Side 1 af 14 Contents Contents... 2 Introduction... 3 Pin assignment of the terminal box connector for customer... 4 Pin assignment of
Læs mereapplies equally to HRT and tibolone this should be made clear by replacing HRT with HRT or tibolone in the tibolone SmPC.
Annex I English wording to be implemented SmPC The texts of the 3 rd revision of the Core SPC for HRT products, as published on the CMD(h) website, should be included in the SmPC. Where a statement in
Læs mereTork Aftørringspapir Standard, M1. Fordel. Produktspecifikation. Beskrivelse. Farve: Hvid Type: Mini Centerfeed Rulle
Fordel Allround aftørringspapir Ideel til rengøring af glas Velegnet til håndaftørring Godkendt til kontakt med fødevarer Tork Easy Handling, emballage som er let at løfte, åbne og folde sammen, når du
Læs mereSpecial VFR. - ved flyvning til mindre flyveplads uden tårnkontrol som ligger indenfor en kontrolzone
Special VFR - ved flyvning til mindre flyveplads uden tårnkontrol som ligger indenfor en kontrolzone SERA.5005 Visual flight rules (a) Except when operating as a special VFR flight, VFR flights shall be
Læs mereThe test can be performed on the following devices. In addition, the required cuvette and the absorption range of the photometer are indicated.
Formaldehyde 10 M. L 1.00-5.00 mg/l HCHO H 2 SO 4 / Chromotropic acid 175 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption
Læs mereKvant Eksamen December 2010 3 timer med hjælpemidler. 1 Hvad er en continuous variable? Giv 2 illustrationer.
Kvant Eksamen December 2010 3 timer med hjælpemidler 1 Hvad er en continuous variable? Giv 2 illustrationer. What is a continuous variable? Give two illustrations. 2 Hvorfor kan man bedre drage konklusioner
Læs mereStatistical information form the Danish EPC database - use for the building stock model in Denmark
Statistical information form the Danish EPC database - use for the building stock model in Denmark Kim B. Wittchen Danish Building Research Institute, SBi AALBORG UNIVERSITY Certification of buildings
Læs mereProject Step 7. Behavioral modeling of a dual ported register set. 1/8/ L11 Project Step 5 Copyright Joanne DeGroat, ECE, OSU 1
Project Step 7 Behavioral modeling of a dual ported register set. Copyright 2006 - Joanne DeGroat, ECE, OSU 1 The register set Register set specifications 16 dual ported registers each with 16- bit words
Læs mereSikkerhed & Revision 2013
Sikkerhed & Revision 2013 Samarbejde mellem intern revisor og ekstern revisor - og ISA 610 v/ Dorthe Tolborg Regional Chief Auditor, Codan Group og formand for IIA DK RSA REPRESENTATION WORLD WIDE 300
Læs mereSTUDIEOPHOLD I BANGKOK FASE 2 - INFORMATION, VEJLEDNING OG DOKUMENTER
Find vejen frem VIA University College STUDIEOPHOLD I BANGKOK FASE 2 - INFORMATION, VEJLEDNING OG DOKUMENTER Find vejen frem VIA University College FASE 2 FASE 2 - INFORMATION, VEJLEDNING OG DOKUMENTER
Læs merePossibilities for Reuse of Calcium Carbonate Pellets from Drinking Water Softening
Possibilities for Reuse of Calcium Carbonate Pellets from Drinking Water Softening Camilla Tang, PhD student Laure Lopato (HOFOR), Sally Nyberg Kornholt (HOFOR) & Hans-Jørgen Albrechtsen (DTU) Danish Water
Læs mereAgenda. The need to embrace our complex health care system and learning to do so. Christian von Plessen Contributors to healthcare services in Denmark
Agenda The need to embrace our complex health care system and learning to do so. Christian von Plessen Contributors to healthcare services in Denmark Colitis and Crohn s association Denmark. Charlotte
Læs mereParticle-based T-Spline Level Set Evolution for 3D Object Reconstruction with Range and Volume Constraints
Particle-based T-Spline Level Set for 3D Object Reconstruction with Range and Volume Constraints Robert Feichtinger (joint work with Huaiping Yang, Bert Jüttler) Institute of Applied Geometry, JKU Linz
Læs mereWander TDEV Measurements for Inexpensive Oscillator
Wander TDEV Measurements for Inexpensive Oscillator Lee Cosart Symmetricom Lcosart@symmetricom.com Geoffrey M. Garner SAMSUNG Electronics (Consultant) gmgarner@comcast.net IEEE 802.1 AVB TG 2009.11.02
Læs mereReexam questions in Statistics and Evidence-based medicine, august sem. Medis/Medicin, Modul 2.4.
Reexam questions in Statistics and Evidence-based medicine, august 2013 2. sem. Medis/Medicin, Modul 2.4. Statistics : ESSAY-TYPE QUESTION 1. Intelligence tests are constructed such that the average score
Læs mereVina Nguyen HSSP July 13, 2008
Vina Nguyen HSSP July 13, 2008 1 What does it mean if sets A, B, C are a partition of set D? 2 How do you calculate P(A B) using the formula for conditional probability? 3 What is the difference between
Læs mereConstant Terminal Voltage. Industry Workshop 1 st November 2013
Constant Terminal Voltage Industry Workshop 1 st November 2013 Covering; Reactive Power & Voltage Requirements for Synchronous Generators and how the requirements are delivered Other countries - A different
Læs mereFokuseret rengøring på sengestuer
Fokuseret rengøring på sengestuer Et clean BED Projekt 2014 2015 Rie Mikkelsen, hygiejnesygeplejerske Indhold clean BED projektet Hypotese Metode Resultater Konklusion Opsummering Referencer clean BED
Læs mereSB ShooeBox. SB Introduction / Indledning 03-03 SB Inspiration combinations/inspirationsopstillinger 04-06 SB Functionality/Funktion 07-07
ShooeBox 02 SB ShooeBox SB Introduction / Indledning 03-03 SB Inspiration combinations/inspirationsopstillinger 04-06 SB Functionality/Funktion 07-07 Introduction Shooebox is a Danish design, designed
Læs merewww.vandibyer.dk IP01: Permeable befæstelser
www.vandibyer.dk IP01: Permeable befæstelser Senior consultant /PhD-fellow Jan Støvring jls@ign.ku.dk Landscape architecture and Planning Project partners: IBF, NCC, BYGGROS, Stenrand Grusgrav, MJK Elektronik,
Læs mereLearnings from the implementation of Epic
Learnings from the implementation of Epic Appendix Picture from Region H (2016) A thesis report by: Oliver Metcalf-Rinaldo, oliv@itu.dk Stephan Mosko Jensen, smos@itu.dk Appendix - Table of content Appendix
Læs merel i n d a b presentation CMD 07 Business area Ventilation
l i n d a b presentation CMD 07 Business area Ventilation 1 Ventilation Sales LTM June 4 097 MSEK EBIT LTM June 449 MSEK 11,0% Two Divisions ADS 85% Comfort 15% YTD June 07 % Sales +31 Nordic countries
Læs mereSlot diffusers. Slot diffusers LD-17, LD-18
LD-17, LD-18 Application LD-17 and LD-18 are designed for supply of cold or warm air in rooms with a height between. m and 4 m. They allow easy setting of air deflectors for different modes of operation
Læs mereUser Manual for LTC IGNOU
User Manual for LTC IGNOU 1 LTC (Leave Travel Concession) Navigation: Portal Launch HCM Application Self Service LTC Self Service 1. LTC Advance/Intimation Navigation: Launch HCM Application Self Service
Læs mereThe use of instrumented gait analysis in interdisciplinary interventions for children with cerebral palsy
PhD thesis The use of instrumented gait analysis in interdisciplinary interventions for children with cerebral palsy Helle Mätzke Rasmussen Department of Clinical Research Faculty of Health Sciences University
Læs mereMontering & Vedligeholdelse/Installation & Maintenance Instructions
Montering & Vedligeholdelse/Installation & Maintenance Instructions Se nyheder og meget mere på wallume.dk For news and more information, visit wallume.dk Tillykke med din Wallume One Med den har du fået
Læs mereCentral Statistical Agency.
Central Statistical Agency www.csa.gov.et 1 Outline Introduction Characteristics of Construction Aim of the Survey Methodology Result Conclusion 2 Introduction Meaning of Construction Construction may
Læs mereSikkerhedsvejledning
11-01-2018 2 Sikkerhedsvejledning VIGTIGT! Venligst læs disse instruktioner inden sengen samles og tages i brug Tjek at alle dele og komponenter er til stede som angivet i vejledningen Fjern alle beslagsdele
Læs mereVarenr.: 553925 90 højre 553926 90 venstre 554027 90º højre med coating 554028 90º venstre med coating
DK GH Skiftespor Varenr.: 55395 90 højre 55396 90 venstre 55407 90º højre med coating 55408 90º venstre med coating 553991 60º højre 553995 60º venstre 551058 60º højre med coating 551059 60º venstre med
Læs mereNordsjællands Hospital. Workshop 5. Sikker medicinering
Nordsjællands Hospital Workshop 5 Sikker medicinering Titel/beskrivelse (Sidehoved/fod) Navn (Sidehoved/fod) 1 Nordsjællands Hospital Program for i dag Titel/beskrivelse (Sidehoved/fod) Navn (Sidehoved/fod)
Læs mereIntelligent Packaging Solutions
Intelligent Packaging Solutions Hofstätter & Ebbesen A/S Højvangen 19 3060 Espergærde Denmark Tel +45 4912 2122 Fax +45 4912 2199 info@ultraplast.dk www.ultraplast.dk UltraPlast is a hi-tech company specializing
Læs mereCross-Sectorial Collaboration between the Primary Sector, the Secondary Sector and the Research Communities
Cross-Sectorial Collaboration between the Primary Sector, the Secondary Sector and the Research Communities B I R G I T T E M A D S E N, P S Y C H O L O G I S T Agenda Early Discovery How? Skills, framework,
Læs mereElma DT 802D CO2 logger. Dansk/Norsk 2 English 4 EAN:
Elma DT 802D CO2 logger Dansk/Norsk 2 English 4 EAN: 5706445840366 Elma DT_802D CO2 logger 2 Brugervejledning Sikkerhed Læs venligst denne brugermanual grundigt igennem og anvend udelukkende instrumentet
Læs mereSustainable use of pesticides on Danish golf courses
Indsæt nyt billede: Sustainable use of pesticides on Danish golf courses Anita Fjelsted - Danish EPA Ministry of the Environment 27 May 2015 - STERF The Danish Environmental Protection Agency 450 employees
Læs mereInstallation Venligst bemærk, håndpumpen kun må monteres i lodret position.
HP-1. HP-3. HP-. Hand Operated Håndpumpe The HP pump is a single acting hand operated pump made of stainless steel St.1.431 making it particularly applicable within the food industry and in environmentally
Læs mereMetodeafprøvning. Bestemmelse af coliforme bakterier og E. coli i spildevand. Naturstyrelsens Referencelaboratorium. Miljøanalyser
Metodeafprøvning Bestemmelse af coliforme bakterier og E. coli i spildevand Naturstyrelsens Referencelaboratorium for Mikrobiologiske Miljøanalyser Titel: Metodeafprøvning, bestemmelse af coliforme bakterier
Læs mereFinancial Literacy among 5-7 years old children
Financial Literacy among 5-7 years old children -based on a market research survey among the parents in Denmark, Sweden, Norway, Finland, Northern Ireland and Republic of Ireland Page 1 Purpose of the
Læs mereChlorine dioxide 50 T mg/l ClO 2 DPD / Glycine
Chlorine dioxide 50 T 0.05-1 mg/l ClO 2 DPD / Glycine 119 Instrument specific information The test can be performed on the following devices. In addition, the required cuvette and the absorption range
Læs mereVejledning til Sundhedsprocenten og Sundhedstjek
English version below Vejledning til Sundhedsprocenten og Sundhedstjek Udfyld Sundhedsprocenten Sæt mål og lav en handlingsplan Book tid til Sundhedstjek Log ind på www.falckhealthcare.dk/novo Har du problemer
Læs mereEngelsk. Niveau D. De Merkantile Erhvervsuddannelser September Casebaseret eksamen. og
052431_EngelskD 08/09/05 13:29 Side 1 De Merkantile Erhvervsuddannelser September 2005 Side 1 af 4 sider Casebaseret eksamen Engelsk Niveau D www.jysk.dk og www.jysk.com Indhold: Opgave 1 Presentation
Læs mereHA-SPAD MEDIA KIT
HA-SPAD0223185808 2017 MEDIA KIT 147,912 1,064,820 WITH 55.6% MOBILE 3.76 (4.24 MOBILE) Source: Omniture Nov. 16 - Jan 17; Nielsen PRIZM 2015, Compete June 2014 - June 2015 Average USA Today Cars.com
Læs mereThe effects of occupant behaviour on energy consumption in buildings
The effects of occupant behaviour on energy consumption in buildings Rune Vinther Andersen, Ph.D. International Centre for Indoor Environment and Energy Baggrund 40 % af USA's samlede energiforbrug sker
Læs mereNovember hilsner fra NORDJYSKE Medier, Distributionen
Uret er stillet til vintertid, og det betyder, at der nu er mørkt både morgen og aften. Det er vigtigt, at du er synlig i trafikken i vintermørket, og derfor opfordrer vi dig til at bruge din refleksvest,
Læs mereDAY HUNTER KIT ASSEMBLY INSTRUCTIONS
DAY HUNTER KIT ASSEMBLY INSTRUCTIONS Attach the Shoulder Harness to the Frame 1. Pass the Upper Harness Attachment Straps (Diagram 1.) through the attachment points located inboard on the upper portion
Læs mereAktivering af Survey funktionalitet
Surveys i REDCap REDCap gør det muligt at eksponere ét eller flere instrumenter som et survey (spørgeskema) som derefter kan udfyldes direkte af patienten eller forsøgspersonen over internettet. Dette
Læs mereHealthcare Apps. OUH Odense University Hospital & Svendborg Hospital. Kiel, Germany, November 2013 1 05/12/13
Healthcare Apps OUH Odense University Hospital & Svendborg Hospital Kiel, Germany, November 2013 1 05/12/13 Jesper Lakman Senior Consultant Digital InnovaGon (4 employees) IT Department (140 employees)
Læs mereHIV and Drug Prevention in Estonian Prisons
HIV and Drug Prevention in Estonian Prisons Maret Miljan The Division of Social Rehabilitation The Estonian Ministry of Justice Prisons in Estonia (3676 prisoners): Tallinn Prison ( 571 convicted male
Læs mereBANGKOK FASE 2 - VALGFAG INFORMATION, VEJLEDNING OG DOKUMENTER
Find vejen frem VIA University College BANGKOK FASE 2 - VALGFAG INFORMATION, VEJLEDNING OG DOKUMENTER Find vejen frem VIA University College FASE 2 Information, vejledning og dokumenter Information, vejledning
Læs mereBiocide resistance in Salmonella Typhimurium
FACULTY OF HEALTH AND MEDICAL SCIENCES UNIVERSITY OF COPENHAGEN PhD thesis Mette Rørbæk Gantzhorn Biocide resistance in Salmonella Typhimurium Academic advisor: Line Elnif Thomsen This thesis has been
Læs mereStatistik for MPH: oktober Attributable risk, bestemmelse af stikprøvestørrelse (Silva: , )
Statistik for MPH: 7 29. oktober 2015 www.biostat.ku.dk/~pka/mph15 Attributable risk, bestemmelse af stikprøvestørrelse (Silva: 333-365, 381-383) Per Kragh Andersen 1 Fra den 6. uges statistikundervisning:
Læs mereDoodleBUGS (Hands-on)
DoodleBUGS (Hands-on) Simple example: Program: bino_ave_sim_doodle.odc A simulation example Generate a sample from F=(r1+r2)/2 where r1~bin(0.5,200) and r2~bin(0.25,100) Note that E(F)=(100+25)/2=62.5
Læs mereModel Brugsanvisning Instruction manual
Model 11560 Brugsanvisning Instruction manual DK GB CAMPINGTOILET Dele Toiletlåg Toiletsæde Vandpåfyldningsstuds Pumpestempel Låsebeslag 10 liters skylletank Tømmetud Skyderventil Niveauindikator 20 liters
Læs mereIHI Triple Aim i Region Midtjylland
IHI Triple Aim i Region Midtjylland v/jens B. Damgaard, kontorchef Strategi & Planlægning, Nære Sundhedstilbud www.regionmidtjylland.dk IHI?? The Institute for Healthcare Improvement (IHI): At IHI, we
Læs mereBANGKOK FASE 2 -VALGFAG INFORMATION, VEJLEDNING OG DOKUMENTER
Find vejen frem VIA University College BANGKOK FASE 2 -VALGFAG INFORMATION, VEJLEDNING OG DOKUMENTER Find vejen frem VIA University College FASE 2 Information, vejledning og dokumenter Information, vejledning
Læs mereSide 1 af 9. SEPA Direct Debit Betalingsaftaler Vejledning
Side 1 af 9 SEPA Direct Debit Betalingsaftaler Vejledning 23.11.2015 1. Indledning Denne guide kan anvendes af kreditorer, som ønsker at gøre brug af SEPA Direct Debit til opkrævninger i euro. Guiden kan
Læs mereBias og confounding. Søren Kold, overlæge, ph.d., klinisk lektor Aalborg Universitetshospital
Bias og confounding Søren Kold, overlæge, ph.d., klinisk lektor Aalborg Universitetshospital Bias og confounding Kritisk litteraturlæsning Introduktion Øvelse Information 5 min. Gruppearbejde 20 min. Diskussion
Læs mereKnee-extension strength or leg-press power after fast-track total knee arthroplasty: Which is better related to performance-based and selfreported
Knee-extension strength or leg-press power after fast-track total knee arthroplasty: Which is better related to performance-based and selfreported function? Peter K Aalund 1, Kristian Larsen 2,3, Torben
Læs mereHelp / Hjælp
Home page Lisa & Petur www.lisapetur.dk Help / Hjælp Help / Hjælp General The purpose of our Homepage is to allow external access to pictures and videos taken/made by the Gunnarsson family. The Association
Læs mereNyhedsmail, december 2013 (scroll down for English version)
Nyhedsmail, december 2013 (scroll down for English version) Kære Omdeler Julen venter rundt om hjørnet. Og netop julen er årsagen til, at NORDJYSKE Distributions mange omdelere har ekstra travlt med at
Læs mereDendrokronologisk Laboratorium
Dendrokronologisk Laboratorium NNU rapport 5, 1997 STRANDGADE 3A, KØBENHAVN Nationalmuseets Marinarkæologiske Undersøgelser. Indsendt af Christian Lemée. Undersøgt af Aoife Daly og Niels Bonde. NNU j.nr.
Læs mereRødsand laboratoriet et samarbejde mellem KU, Femern & DHI
Rødsand laboratoriet et samarbejde mellem KU, Femern & DHI Ulrik Lumborg DHI Rødsand laboratoriet I 2012 var det erkendt at Rødsand lagune ville være et fokusområde i forbindelse med etableringen af Femernforbindelsen
Læs mereTo the reader: Information regarding this document
To the reader: Information regarding this document All text to be shown to respondents in this study is going to be in Danish. The Danish version of the text (the one, respondents are going to see) appears
Læs mereMidlertidig ændring af vandstand som redskab til vedligeholdelse af lobeliesøer
Lake Skærsø, Denmark Midlertidig ændring af vandstand som redskab til vedligeholdelse af lobeliesøer Anna-Marie Klamt, Kasper Reitzel, Frede Ø. Andersen, Henning S. Jensen CLEAR temadag 24 May 2016 Syddansk
Læs mereframe bracket Ford & Dodge
, Rev 3 02/19 frame bracket 8552005 Ford & Dodge ITEM PART # QTY DESCRIPTION 1 00083 8 NUT,.50NC HEX 2 00084 8 WASHER,.50 LOCK 3 14189-76 2 FRAME BRACKET 4 14194-76 1 411AL FRAME BRACKET PASSENGER SIDE
Læs mereDiamond Core Drilling
Diamond Core Drill ZIZ-250Pro Product Description Diamond core drill is a tool used for drilling holes on reinforced concrete, rocks and fire-proof materials. It features in high-efficiency, smooth and
Læs mereDen nye Eurocode EC Geotenikerdagen Morten S. Rasmussen
Den nye Eurocode EC1997-1 Geotenikerdagen Morten S. Rasmussen UDFORDRINGER VED EC 1997-1 HVAD SKAL VI RUNDE - OPBYGNINGEN AF DE NYE EUROCODES - DE STØRSTE UDFORDRINGER - ER DER NOGET POSITIVT? 2 OPBYGNING
Læs mereElektriske apparater til husholdningsbrug o.l. Sikkerhed Del 1: Generelle krav
Dansk standard Tillæg DS/IEC 60335-1/A1:2013 + Corr 1:2014 1. udgave 2014-05-08 Elektriske apparater til husholdningsbrug o.l. Sikkerhed Del 1: Generelle krav Household and similar electrical appliances
Læs mereESG reporting meeting investors needs
ESG reporting meeting investors needs Carina Ohm Nordic Head of Climate Change and Sustainability Services, EY DIRF dagen, 24 September 2019 Investors have growing focus on ESG EY Investor Survey 2018
Læs mereUNISONIC TECHNOLOGIES CO.,
UNISONIC TECHNOLOGIES CO., 3 TERMINAL 1A NEGATIVE VOLTAGE REGULATOR DESCRIPTION 1 TO-263 The UTC series of three-terminal negative regulators are available in TO-263 package and with several fixed output
Læs mereHudevad P5 Easy Clean
Technical datasheet Key benefits Rengøring Unique tilting front Radiator front tilts away from the wall for easy access to cleaning the back of the radiator Classic Hudevad design Trademark Hudevad design
Læs mereIBM Network Station Manager. esuite 1.5 / NSM Integration. IBM Network Computer Division. tdc - 02/08/99 lotusnsm.prz Page 1
IBM Network Station Manager esuite 1.5 / NSM Integration IBM Network Computer Division tdc - 02/08/99 lotusnsm.prz Page 1 New esuite Settings in NSM The Lotus esuite Workplace administration option is
Læs mereMandara. PebbleCreek. Tradition Series. 1,884 sq. ft robson.com. Exterior Design A. Exterior Design B.
Mandara 1,884 sq. ft. Tradition Series Exterior Design A Exterior Design B Exterior Design C Exterior Design D 623.935.6700 robson.com Tradition OPTIONS Series Exterior Design A w/opt. Golf Cart Garage
Læs mereVEDLIGEHOLDELSE AF STENMØBLER
DK VEDLIGEHOLDELSE AF STENMØBLER DK / FORORD Sten er et af naturens smukkeste materialer og er skabt gennem tusindvis af år. Derfor er to møbler af sten aldrig helt ens. Disse forskelligheder er skabt
Læs mereEconomy The cost minimization has finally started paying off, and will be used as a motivational factor in 2015.
English resume The English resume is a short version of the minutes from the board meeting in order to give our international members an insight to the meetings. The English resume is not as detailed as
Læs mereMotorway effects on local population and labor market
Motorway effects on local population and labor market Per Homann Jespersen Associate Professor, Roskilde University Jean P. Endres phd student, Roskilde University Trafikdage 23-08-16 Motorways and the
Læs mereTender Package 10 pre-description
Tender Package 10 pre-description August 2012 Banedanmark Amerika Plads 15 2100 København Ø www.banedanmark.dk Content Page TENDER PACKAGE TP 10 VESTVOLDEN 4 General description of TP 10 4 List of 5 List
Læs mereCOPYRIGHT Danish Standards. NOT FOR COMMERCIAL USE OR REPRODUCTION. DS/ISO :2012
Dansk standard DS/ISO 11269-2 2. udgave 2012-03-20 Jordkvalitet Bestemmelse af forurenende stoffers indflydelse på jordflora Del 2: Forurenet jords indflydelse på højere planters fremspiring og tidlige
Læs mereNYC Reservoir System Current Operations December 14, 2010
NYC Reservoir System Current Operations December 14, 2010 Outline System Status Storm Events Impacts to Reservoirs NYC Response Current Operations Plan moving forward System Status Consumption NYC + upstate
Læs mereSEPA Direct Debit. Mandat Vejledning 2013.03.15. Nets Lautrupbjerg 10 DK-2750 Ballerup
SEPA Direct Debit Mandat Vejledning 2013.03.15 Nets Lautrupbjerg 10 DK-2750 Ballerup Indholdsfortegnelse 1. Indledning... 3 1.1 Tilknyttet dokumentation... 3 1.2 Kontakt til Nets... 3 2. Krav til SEPA
Læs mereDendrokronologisk Laboratorium
Dendrokronologisk Laboratorium NNU rapport 14, 2001 ROAGER KIRKE, TØNDER AMT Nationalmuseet og Den Antikvariske Samling i Ribe. Undersøgt af Orla Hylleberg Eriksen. NNU j.nr. A5712 Foto: P. Kristiansen,
Læs mereTimetable will be aviable after sep. 5. when the sing up ends. Provicius timetable on the next sites.
English Information about the race. Practise Friday oct. 9 from 12.00 to 23.00 Saturday oct. 10. door open at 8.00 to breakfast/coffee Both days it will be possible to buy food and drinks in the racecenter.
Læs mere